CONTROLS AND INDICATORS
POWER SUPPLY
POWER
CONTROL
VOLTAGE
MAIN POWER
LINE VOLTAGE
+250 VDC
+345 VDC
+24 VDC
-20 VDC
PNEUMATICS
+9 VDC
-9 VDC
+20 VDC
PNEU TEMP
ELEC TEMP
POWER SUPPLY
5 PSI
INCREASE
30 PSI
INCREASE
VACUUM
60 PSI
30 PSI
5 PSI
150
0
050
100
200
250
INIT VOLTAGE
OFF
OFF
OFF
OFF
Hialeah, Florida USA
COULTER ELECTRONICS, INC.
300
OFF
ON
The VOLTAGE LAMPS
light when the power
supply voltages are
out of preset tolerance,
or when the corresponding
fuse is blown.
The LINE VOLTAGE
METER indicates that
voltage is present.
The CONTROL VOLTAGE
LAMP, when lit indicates
sufficient voltage is present
at the Diluter.
The PNEUMATICS LAMP
lights when the 60 PSI
gauge is off or low
(Vacuum<17; pressure <45).
During the cycle, a momentary
vacuum drop below 17" Hg
does not constitute a
system failure.
The PNEU TEMP (Pneumatic
temperature) lamp lights when
the pneumatic assembly is
too hot: the pneumatics
subsystem turns off.
The ELEC TEMP (Electronic
temperature) lamp lights when
the electronic assembly is
too hot: the electronics
assembly and the pneumatics
subsystem turn off.
Turn the 5 PSI KNOB to
regulate the low pressure
used for reagent level
sensors, mixing bubbles
and needle vent.
The VACUUM TRAP BOTTLE
protects the vacuum pump
from liquid. Liquid in this
bottle indicates a problem.
WARNING: May contain
biohazardous material.
Turn the 30 PSI KNOB
to regulate the high
pressure used to move
components and move
fluids through the Diluter.
The MAIN POWER CIRCUIT
BREAKER turns power to the
System on and off.
7188018A
The PNEUMATIC GAUGES (4)
indicate the amount of
pressure and vacuum present.
8
PN 4237188A (December 1993)
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......