CYCLING SAMPLES
c.
DILUTER
ID
Enter 3 - 11 digits.
1
1
1
ENTER
d.
Immerse the aspirator tip in the sample
tube.
e.
Press and release the activator.
7188052A
IMPORTANT
Do not remove the sample tube from the aspirator tip until the STATUS message on the Analyzer CRT
changes from
INTRO SAMPLE to WIPE, and there is a beep.
WARNING
The aspirator tip retracts when the
PRIME message appears in the STATUS field on the Analyzer CRT;
be sure to remove the sample tube and be clear of the aspirator tip before the
PRIME message
appears.
f.
Remove the tube when
•
you hear the beep and
•
the Analyzer CRT STATUS field
displays
WIPE
.
PN 4237188A (December 1993)
63