CHECK STARTUP RESULTS
3.
If the background results or Hgb readings do
not meet the criteria listed in steps 1 and 2,
repeat the Background test at the Analyzer
CRT. The Background test includes Hgb
voltage check and diff test.
Press in sequence the key next to
ANALYZER
MAIN
ANALYZER FUNCTIONS
STARTUP TESTS
BACKGROUND TEST
If results are still not acceptable, call Coulter
Customer Operations. Return to the System
Run display.
4.
Check the ramp and precision tests. If any
test value exceeds the reference value by 1%
or more, the test value appears in red on the
DMS screen and is flagged with an H (high)
or L (low) flag. If a flag occurs:
a.
Select
DMS
Special Functions
Set Up
System Set Up
Electronic Reference
to see if the values are correct.
b.
See if your laboratory has calibrated the
instrument recently.
PN 4237188A (December 1993)
17