CONTROLS AND INDICATORS
DILUTER KEYPAD
DRAIN
RINSE
CLEAR
APERT
START/
CONT
POWER
OFF
START
UP
SHUT
DOWN
ALARM
RESET
PREMIX
STOP
7
8
9
5
6
1
2
3
CE
0
F
ID
ENTER
7188020A
PRIME
APERT
POWER
ON
4
Turns on power to the Analyzer and Diluter.
POWER
ON
PRIME
APERT
When pneumatics are OFF:
•
Activates pneumatics.
•
Backwashes both modes.
•
Drains and rinses baths.
•
Primes Hgb and lyse pumps.
•
Applies 6" vacuum to apertures to
remove bubbles from lines.
•
Turns on aperture current (burn
cycle).
•
Rotates and rinses mixing
chamber.
When pneumatics are ON:
•
Drains and rinses baths.
•
Backwashes both modes.
•
Applies 6" vacuum to apertures
to remove bubbles from lines.
•
Rotates and rinses mixing
chamber.
Press during cycling in the Primary mode to increase light intensity for a better view of
the apertures.
Press after
when you are entering the 5C
®
and Retic-C cell control numbers for
ID
the Secondary mode.
Empties the RBC bath and vacuum isolator chambers into the waste chamber.
DRAIN
Empties the WBC bath through the Hgb cuvette into the waste chamber.
Fills baths with diluent.
RINSE
Empties the waste chamber.
4
PN 4237188A (December 1993)
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......