SAMPLE ANALYSIS
MODES OF OPERATION
Sample Modes (Diluter Aspiration)
Sample mode determines which Diluter aspiration method is used.
Primary Mode
•
Closed vial
•
Automated
•
Requires 250 µL of whole blood for a complete CBC or CBC with WBC differential. There must
be a minimum of 1 mL of sample with the proper proportion of blood to anticoagulant in the
tube.
Secondary Mode
•
Open vial
•
Requires only 150 µL of whole blood for a complete CBC or CBC with WBC differential.
•
Use for samples with very low hemoglobin results when Primary mode run produces a partial
|
aspiration. (Blood detectors do not recognize these samples as blood.)
|
•
Use for Reticulocyte samples. Aspirates approximately 2 mL of prepared Retic sample. 50 µL of
whole blood required to prepare the sample.
•
Use for LATRON control and primer, with specific Diluter functions per Test mode.
Test Modes (Analyzer Configurations)
Test mode, set at the Analyzer CRT, determines how the data is processed by the STKS.
CBC Mode
Performs basic CBC on whole-blood samples, in either Primary or Secondary mode.
PN 4237188B (March 1995)
47
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......