DAILY QUALITY CONTROL
DAILY QUALITY CONTROL
IMPORTANT
When 100 total runs are stored in a control file,
FILE IS FULL appears on the error line.
LATRON
LATRON verifies flow cell alignment, gains and CVs for flow cell volume, conductivity and scatter.
You can use LATRON to monitor Diff and Retic modes in three different ways. Each method uses a
different Diluter function. Each requires approximately 1.5 mL of LATRON.
Functions for using LATRON:
•
Diff only
- F55
•
Retic only
- F56
•
Diff + Retic
- F57
Coulter recommends using F57 for monitoring Diff and Retics with the LATRON.
The following procedure is for Diff + Retic. If you are using LATRON for Diff only or Retics only,
choose the appropriate function. For most efficient reagent use, use the function for what you are
going to run.
1.
Make sure the COULTER LATRON primer
and control are within the correct
temperature range before use.
˚F
˚C
30
86
18
64
5935019C
2.
NEW
LOT NO.
?
NO
YES
CONTROLS,
SET UP
CONTINUE
7188035A
PN 4237188A (December 1993)
19
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......