DAILY QUALITY CONTROL
7.
Select
DMS
Controls
Control Run
Press
Sample Analysis
Controls
Start Up
Special Functions
Control Run
Review or Report
Graphs
Control Batch
COULTER
≡≡≡
STKS
CBC/Diff/Retic
(931251) COPYRIGHT 1989-1993 COULTER CORPORATION, ALL RIGHTS RESERVED
F1-Help
F3-Error Log
F5-Other
F9-Main menu
F10-Previous menu
05/20/93 10:33
OPR
DMS
↑
PR
↓
TK
↓
HC
↓
DB
↑
XB
↓
WL
↑
HWL
↓
QC
↑
73
8.
Press
File
if the file you want is not on
the screen.
9.
In the Select file window, move the cursor to
the file you want.
Press
.
RETIC CONTROL RUN
FILE:
1
LOT: 311000
OPR: abc
SHIFT : 0
LEVEL: Level I
EXP DATE: 12/31/93
RUN:
6
DATE:
Select File:
Level
Lot#
Shift
Level
Lot#
Shift
Ref RB Level I
311000
0
NOT SETUP
Level II
421000
0
NOT SETUP
Level III
531000
0
NOT SETUP
LATRON D+R
107366
NOT SETUP
RET%
Normal
884300
0
NOT SETUP
Abnormal I
874300
0
NOT SETUP
MRV
Abnormal II
864300
0
NOT SETUP
NOT SETUP
NOT SETUP
MI
NOT SETUP
NOT SETUP
NOT SETUP
NOT SETUP
NOT SETUP
NOT SETUP
NOT SETUP
NOT SETUP
NOT SETUP
NOT SETUP
RET#
NOT SETUP
NOT SETUP
NOT SETUP
NOT SETUP
F1-Help
F2-File
F4-Print
F6-Graph
F8-Delete
F9-Menu
F12-Graphics
10/07/93 08:09
OPR
DMS
↑
PR
↓
TK
↓
HC
↑
DB
↑
XB
↑
WL
↓
HWL
↓
QC
↑
48
The selected CONTROL RUN file appears.
The box in the lower right corner of the
screen tells you which mode you are in and
which function to use for this mode.
In a Diff + Retic Latex file,
and
toggle between DIFF and RETIC displays.
DIFF LATEX CONTROL RUN
FILE:
3
LATRON
RUN:
0
EXP DATE: 03/21/94
LOT: 107359
Mean Channel
RESULTS
ASSAY
DIFF
RANGE
+/-
Volume (V)
27.0
2.0
Conductivity (C)
27.0
2.0
Scatter (S)
00.0
5.0
% CV
RESULTS
EXPECTED CV
<
Volume (V)
7.0
Diff+Retic
Conductivity (C)
10.0
REMEMBER TO RUN
Scatter (S)
9.0
IN FUNCTION 57
F1-Help F2-File F3-Primer F4-Prt F5-Histo F6-Graph F7-Next F8-Del F9-Menu
↔
Next
10/03/93 10 41
OPR
DMS
↑
PR
↑
TK
↓
HC
↓
DB
↑
XB
↑
WL
↓
HWL
↓
QC
↑
10. Press
Primer
.
IMPORTANT
Use the correct Diluter function when you run the
primer.
CAUTION
Use of F55, F56 or F57 for anything other than
the
LATRON
procedure
can
damage
the
instrument.
DIFF LATEX CONTROL RUN
FILE:
1
LATRON D+R
RUN:
5
EXP DATE
PRIMER RUN
LOT: 107999
10/10/93
07:09:28
DATE:10/11/93
TIME:07:24:11
Mean Channel
RESULTS
ASSAY
DIFF
COUNT
Diff
6
Volume (V)
27.0
Retic
24
Conductivity (C)
27.0
Scatter (S)
90.0
5.0
% CV
RESULTS
EXPECTED CV
<
Volume (V)
7.0
Diff+Retic
Conductivity (C)
10.0
REMEMBER TO RUN
|
Scatter (S)
9.0
IN FUNCTION 57
|
F1-Help
F3-Latex
F4-Print
F8-Delete
F9-Menu
10/11/93 07 24
OPR
DMS
↑
PR
↑
TK
↓
HC
↑
DB
↑
XB
↑
WL
↓
HWL
↓
QC
↑
49
PN 4237188A (December 1993)
21
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......