STARTUP
AT THE POWER SUPPLY
CAUTION
Do not operate the instrument if any indicator is outside the following limits.
The pneumatic subsystem has an automatic shutoff feature to prolong the life of the compressor and
vacuum pump. If the Diluter is not cycled for about one hour, the pneumatic subsystem turns off.
Some error messages can inhibit the time out function.
If the pneumatics are off,
DILUTER
PRIME
APERT
1.
Verify that:
•
LINE VOLTAGE meter is between
105 V and 125 V (or 210 V - 250 V).
•
CONTROL VOLTAGE lamp is lit.
•
All MONITOR LAMPS are unlit.
WARNING
The contents of the vacuum trap bottle may
include residual biological material and must be
handled with care. Avoid skin contact and clean
up spills immediately. Dispose of the liquid
following acceptable laboratory procedures.
•
Vacuum trap is free of liquid.
If there is liquid, see the Vacuum Trap
section in Special Procedures and
Troubleshooting, PN 4237187.
LINE
VOLTAGE
METER
VACUUM
TRAP
MONITOR
LAMPS
5 PSI
REGULATOR
5 PSI
GAUGE
VACUUM
GAUGE
60 PSI
GAUGE
30 PSI
GAUGE
30 PSI
REGULATOR
CONTROL
VOLTAGE
LAMP
7188011A
12
PN 4237188A (December 1993)
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......