DAILY QUALITY CONTROL
4.
When the cycle is finished,
Other
.
DMS
Prime
to turn Prime off.
.
5.
DMS
Select
Controls
Control Run
and press
.
Select the CBC control file for
• the level, and
• the lot number
you are going to use.
CBC CONTROL RUN
FILE:
9
LOT: 085300
OPR: 203
SHIFT : 0
LEVEL: 4C Abn. High
EXP DATE: 22/12/93
RUN:
6
13/10/19
09:34
RESULTS
ASSAY
DIFF
/-
WBC
19.2
0.9
RBC
5.23
0.14
HGB
16.6
0.4
HCT
47.8
2.5
MCV
91.4
3.0
MCH
31.7
1.5
MCHC
34.7
2.2
RDW
13.8
2.0
PLT
389
40
MPV
10.4
2.0
REMEMBER TO
DISABLE THE DIFF
F1-Help
F2-File
F4-Print
F6-Graph
F8-Delete
F9-Menu
20/10/93 09:07
OPR
DMS
↑
PR
↓
TK
↓
HC
↓
DB
↑
XB
↑
WL
↑
HWL
↓
QC
↑
46
2
6.
Prepare the 4C PLUS cell control (or
equivalent) according to directions on the
package insert.
7.
Put the control tubes
•
LOW
•
NORMAL
•
HIGH
preferably in that order from left to right in a
cassette.
7188064A
PN 4237188A (December 1993)
43
Содержание STKS
Страница 16: ...CONTROLS AND INDICATORS 6 PN 4237188A December 1993...
Страница 84: ...SAMPLE ANALYSIS 74 PN 4237188A December 1993...
Страница 98: ...DATA ANALYSIS 88 PN 4237188A December 1993...
Страница 124: ...DMS BASICS 114 PN 4237188A December 1993...
Страница 130: ...RUN SAMPLES DISPLAY 120 PN 4237188A December 1993...
Страница 168: ...DATA BASE 158 PN 4237188A December 1993...
Страница 204: ...APPENDIX A 194 PN 4237188A December 1993...
Страница 216: ...APPENDIX C 206 PN 4237188A December 1993...
Страница 228: ...APPENDIX D 218 PN 4237188B March 1995...
Страница 230: ...REFERENCES 220 PN 4237188B March 1995...
Страница 236: ......