Manual Operation
3-41
2.
Use the numeric keys (
0
~
9
)
or “Rotary” (
) knob to set the value.
3.
Press “
ENTER
” to confirm.
4.
Press “
EXIT
” to return to the MAIN PAGE.
This function sets the protection point for Over Voltage. Once the output voltage exceeds
the range, it will turn off the output that is OUTPUT = OFF to protect the unit under test.
Table 3-2 shows the voltage range of OVP.
Table 3-2 OVP Range
Model
Min. OVP (V)
Max. OVP (V)
62xxxH-xxxxS
0
1.10 x Vo_MAX
When OVP occurs the main page will prompt a protection message as Figure 3-70 shows:
V = 6 0. 0 0 V I = 1 0. 0 0 _ A O F F
0 . 0 0 0 0 V 0 . 0 0 0 0 A
O V P 0 . 0 W
Figure 3-70
3.3.5.2
OCP Protection
1.
Use “
”, “
” keys to move the cursor to the column to be set as Figure 3-71
shows.
[ P R O T E C T I O N ]
O V P
= 6 0 0 . 0 V
O C P
= 8 . 9 2 _ A
O P P
= 5 2 5 0 . 0 W
R E M O T E I N H I B I T
= D I S A B L E P U L L = H I G H
S A F E
T Y I N T . L O C K
= D I S A B L E P U L L = H I G H
E X T O N / O F F = D I S A B L E P U L L = H I G H
▼
Figure 3-71
2.
Use the numeric (
0
~
9
) keys or “Rotary” (
) knob to set the value.
3.
Press “
ENTER
” to confirm.
4.
Press “
EXIT
” to return to the MAIN PAGE.