Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
3-30
3.
Press “
ENTER
to confirm, it will skip to the series/parallel MASTER page automatically as
Figure 3-45 shows.
∑ V = 6 0. 0 0 _ V I = 1 0. 0 0 A M S T O F F
0 . 0 0 0 0 V 0 . 0 0 0 0 A
0 . 0 W
Figure 3-45
4. The SLAVE page will skip to Figure 3-46 automatically.
O N L I N E
S L A V E 1
Figure 3-46
5. Start the series/parallel usage.
CAUTION
1.
Communication error will occur if the SLAVE settings are the same
(such as SLAVE 1 & SLAVE 1). The MAIN PAGE of MASTER will
show as Figure 3-47. When this type of error occurs, exit the
series/parallel operation first and then change the SLAVE setting to
resume the operation.
S L A V E 1
M E A S . E R R
Figure 3-47
2.
Once the series/parallel is set, the settings can be saved. After all
machines are powered off, turn on the SLAVE first and MASTER
the last, it will set series/ parallel operation automatically.