Programmable DC Power Supply (with Solar Array Simulation) 62000H Series
Operating & Programming Manual
3-76
2.
In the CALIBRATION page, press
“
1
” or turn the “Rotary” (
) knob to set
CHOICE=1.
3.
Press
“
ENTER
” to confirm entering into Voltage Calibration as Figure 3-137 shown
below.
[ V O L T A G E C A L I B R A T I O N ]
R E M O V E E X T E R N A L L O A D A N D P R E S S [ E N T E R ] _
S E T T I N G O U T P U T V O L T A G E
1 0 . 0 0
V
A C T U A L
O U T P U T V O L T A G E = 0 . 0 0 0 0
V
S E T T I N G O U T P U T V O L T A G E
2 9
. 0 0
V
A C T U A L
O U T P U T V O L T A G E = 0 . 0 0 0 0
V
Figure 3-137
4.
When in Voltage Calibration page, press
“
ENTER
” to confirm.
5.
First perform low voltage range calibration. The device will output 10.00V and the
cursor will stop at position [1] as shown in Figure 3-138. Enter the voltage read by
DVM to position [1] and press
“
ENTER
” to confirm it.
6.
Next, press
“
” to do the low voltage range second point calibration. The device
will output 29.00V and the cursor will stop at position [2] as shown in Figure 3-138.
Enter the voltage read by DVM to position [2] and press
“
ENTER
” to confirm it.
[
V O L T A G E C A L I B R A T I O N
]
R E M O V E E X T E R N A L L O A D A N D P R E S S[ E N T E R ]
S E T T I N G O U T P U T V O L T A G E 1 0 . 0 0
V
A C T U A L
O U T P U T V O L T A G E= 9 . 9 9 9 8 _
V
S E T T I N G O U T P U T V O L T A G E 2 9 . 0 0
V
A C T U A L
O U T P U T V O L T A G E= 2 8 .
V
[ 1 ]
[ 2 ]
9 9 8 2
Figure 3-138
7.
Press
“
” again to start the high voltage range calibration. The device will output
38.00V and the cursor will stop at position [3] as shown in Figure 3-139. Enter the
voltage read by DVM to position [3] and press
“
ENTER
” to confirm it.
8.
Press
“
” one more time to do the high voltage range calibration. The device
output 132.00V and the cursor will stop at position [4] as shown in Figure 3-139. Enter
the voltage read by DVM to position [4] and press
“
ENTER
” to confirm it.