Remote Operation
5-39
Example:
PROG:SEQ:CURR 40.5
Query Syntax:
PROG:SEQ:CURR?
Return Parameter: <NRf+>
Query Example: PROG:SEQ:CURR?
Return Example: 4.01
12. PROGram:SEQuence:CURRent:SLEW
Description:
It sets sequence for output voltage slew rate.
Syntax:
PROGram:SEQuence:CURRent:SLEW <NRf+>
Example:
PROG:SEQ:CURR:SLEW 10
Query Syntax:
PROG:SEQ:CURR:SLEW?
Return Parameter: <NR1>
Query Example: PROG:SEQ:CURR:SLEW?
Return Example: 1.00
13. PROGram:SEQuence:CURRent:SLEWINF
Description:
It sets the slewrate of sequence current output to INF.
Syntax:
PROGram:SEQuence:CURRent:SLEWINF ENABLE
PROGram:SEQuence:CURRent:SLEWINF DISABLE
Parameter:
ENABLE/DISABLE
Example:
PROGram:SEQuence:CURRent:SLEWINF ENABLE sets the
Slewrate to INF
PROGram:SEQuence:CURRent:SLEWINF DISABLE releases the
Slewrate INF and return to 1A/ms
Query Syntax:
PROGram:SEQuence:CURRent:SLEW?
Return Parameter: INF. Or <NRf+>[Unit Amp]
Query Example: PROGram:SEQuence:CURRent:SLEW? It returns the settings.
Return Example: INF.
14. PROGram:SEQuence:TIME
Description:
It sets the sequence for the duration of time.
Syntax:
PROGram:SEQuence:TIME <NRf+>
Parameter:
0.005~15000, 0 (0 means end)
Example:
PROG:SEQ:TIME 10
Query Syntax:
PROG:SEQ:TIME?
Return Parameter: <NR f+>
Query Example: PROG:SEQ:TIME?
Return Example: 1.01
15. PROGram:CLEAR
Description:
It clears the sequence.
Syntax:
PROGram:CLEAR
Example:
PROG:CLEAR
16. PROGram:ADD
Description:
It adds a sequence.
Syntax:
PROGram:ADD <NR1>
Parameter:
1~100 (based on the remaining SEQUENCE no. for configuration)
Example:
PROG:ADD
Query Syntax:
PROGram:ADD?
Return Parameter: <NR1>
Query Example:
PROGram:ADD?
Return Example:
85
– it indicates the remaining no. is 85.