SKEMA 6
-
OPERATING INSTRUCTION
153
GB
153
EN
8.
Accessories
8.1.
Operating lamp
The operating lamp comes in 2 models:
Lamp with halogen light source – model VENUS PLUS.
The instructions for use and maintenance come with the lamp.
2
Lamp with LED light source – model VENUS PLUS-L.
The instructions for use and maintenance are given in Paragraph
8.1.1.
NOTE: during the automatic movements of the dental chair, the
lamp automatically turns off to prevent blinding the patient.
8.1.1.
VENUS PLUS-L OPERATING LAMP
Technical specifi cations.
• Operation: continuous.
• Power supply voltage: 18 V~ 50/60 Hz.
• Power absorbed: 15 VA.
• Light spot dimensions: 65 mm x 165 mm.
• Lux: 50,000 (max).
• Colour temperature: 5,000 °K (+/- 5%).
• Maximum temperature on accessible surfaces (during normal use): < 50°C.
• Protection against electric risks: Class II type B.
• Light source: 2 multichip LEDs.
• De
vice classifi cation in accordance with EN 62471: 2008 for PHOTOBIO-
LOGICAL SAFETY: Risk Group 1.
Safety warnings.
WARNING!
• The lamp must be installed by authorised staff.
• The lamp may only be used by authorised and adequately trained
staff (dentists and paramedics).
• Do not apply any weight on any point of the lamp and act on it only
with the force necessary to move it.
• An operator must always be present when the lamp is on, in parti-
cular, it should not be left unattended in the presence of children/
incompetent persons or in general persons not authorised for use.
• Do not carry out any maintenance operation on the lamp when the
power is on. Cut the power before carrying out any operation on the
lamp.
• The lamp is not suitable for use in the presence of a mixture of
fl ammable anaesthetic gas with oxygen or nitrous oxide.
• The cleaning operations must be carried out with the lamp off and
at room temperature. Do not spray detergent liquid directly onto
parts of the optical unit.
• The lamp is not protected against penetration of liquids (IPX0).
• Do not apply protections, not even transparent, on the optical unit,
which may obstruct the cooling air intakes.
Description.
a) Light source composed of 2 multichip LEDs.
c) Removable handles sterilizable in an autoclave.
d) Power switch.
e) Luminous intensity adjustment knob.
k) NO TOUCH sensor for turning on the lamp and adjusting the luminous
intensity (optional).
f) Self-balancing
arm.
Turning the lamp on and off.
• To turn on the lamp, fl ick the switch (
d
) upward.
• To turn off the lamp fl ick the switch (
d
) upward again.
NOTE for lamps with NO TOUCH sensor: you can also turn the
lamp on and off by rapidly sweeping your hand across the sensor ( k )
(sensitivity range 5 cm).
If the lamp does not turn on or off when sweeping your hand rapidly across
the sensor, sweep it across again more slowly.