SKEMA 6
-
OPERATING INSTRUCTION
158
GB
158
EN
How to remove the separator’s container
ATTENTION!
Gloves must be worn when carrying out the following operation to
prevent contact with infected material.
• Move the dental chair fully up.
• Open the cover at the side of the hydrogroup
[ a ]
(see paragraph 7.7)
and turn the electrical box
[ b ]
.
• Turn the electrical box
[ b ]
and hydrogen peroxide tank
[ a ]
(if present).
• Completely empty the separator’s container by pressing the timed button
[ c ]
found on the cover.
• If present, remove the valve [ s ] for central systems.
• Turn and raise the container until it is detached from the drain pump (k).
• Detach the container
[ d ]
from the cover
[ f ]
by pulling up the two side
elastic bands
[ e ]
.
• Aft
er the cleaning operations, refi t the vessel [
d
] after lubricating the
O-rings with S1-Protective Lubricant for O-Rings (CEFLA S.C.).
• Put the electrical box
[ b ]
back in place and put the cover
[ a ]
back on
the side of the hydrogroup (see paragraph 7.7).
9.6. Cleaning the turbine return air fi lter
Monthly check the oil container fi lter
[ g ]
present in the turbine’s return
air line.
If necessary, replace the fi lter element (code 97290014).