166
SKEMA 6
-
OPERATING INSTRUCTION
166
GB
166
EN
When needed.
Removable instrument cords
Clean with a suitable disinfectant carefully
following the directions for use provided by
the manufacturer. Spray the product on soft
disposable paper towels.
Do not use acids or harsh products.
See paragraph 5.
Spray ducts
Disinfect water ducts of sprays with manual
cycle (if the S.S.S. system is present).
See paragraph 7.2.1.
Disinfect water ducts of sprays with semi-
automatic cycle (if the AUTOSTERIL system
is present).
See paragraph 7.2.
CATTANI surgical separator
Clean the separator's container, drain valve
and probes.
See paragraph 9.5.
METASYS amalgam separator
Empty the separator bowl.
See documentation enclosed
with the separator.
DURR amalgam separator
Empty the separator bowl.
See documentation enclosed
with the separator.
Operatory light
Clean the front glass and refl ector mirror.
See documentation enclosed
with the equipment.
VENUS PLUS-L LED operating
lamp
Clean the transparent screen and the handles.
See paragraph 8.1.1.
Monitor / Workstation with light
pole
Clean the surfaces.
See documentation enclosed
with the equipment.
Coated surfaces and dental
chair upholstery
Clean with a suitable disinfectant carefully
following the directions for use provided by
the manufacturer. Spray the product on soft
disposable paper towels.
Do not use acids or harsh products.
See paragraph 1.4.
Weekly.
Suction tube holders
Lubricate the o-rings.
See paragraph 9.4.
Monthly.
Turbine return air fi lter
Check the fi lter and replace it if necessary
(code 97290014).
See paragraph 9.7.
H.P.A. fi lter
Sterilize the fi lter cartridge.
See paragraph 8.5.
Yearly.
Dental chair and dental unit
Contact the technical service department for
general inspection.
/