SKEMA 6
-
OPERATING INSTRUCTION
116
GB
116
EN
5.4. Turbine
Connecting the handpiece and changing the chuck.
Refer to the specifi c instructions furnished with the handpiece.
Use.
•
Operating time: work 5 min., rest 5 min.
• Place the instrument in its work position.
• Use the foot control pedal to start the instrument (see paragraph 5.2).
• The cock (
f
) adjusts the water fl ow to the spray.
• The cock (
e
) adjusts the amount of air spray for all the instruments.
ATTENTION
!
The instrument is supplied non-sterile.
Adjusting the turbine’s operating speed.
• Place the instrument in its work position.
• Adjust the turbine’s max. rpm with buttons INCREASE and/or DECRE-
ASE.
NOTE: the following values can be set using the fast selection
buttons:
- button “MAX”: 100% power,
- button “MED”: 50% power,
- button “MIN”: 1% power.
• Use the right multifunction button to select the mode for varying the
turbine’s rpm.
The icon for the selected mode appears at the multifunction button.
a) Icon
: Linear variation, proportional to the degree the foot control
lever is moved.
b) Icon
: ON/OFF mode which sees that the set maximum power is
delivered when the foot control lever is actuated.
NOTE: the set data are automatically saved.
Turning the fi ber optic on/off.
• Set the instrument to its work position.
• Press the button
to turn on the fi ber optic.
NOTE: icon ( A ) appears on the display to signal the fi ber optic
has been turned on.
• Adjust brightness with the “increase” (+) and/or “decrease” (-) buttons.
The set level is automatically saved.
Adjusting brightness of the fi ber optic.
• Hold down the button (at least 2 seconds) to adjust brightness of the fi ber
optic
.
• Use buttons INCREASE and/or DECREASE to adjust the brightness level.
NOTE: a value ranging from 1 to 16 can be set.
The following values can be set using the fast selection buttons:
- button “MAX”: level 16,
- button “MED”: level 8,
- button “MIN”: level 1.
• To confi rm the selected brightness level, press multifunction button ESC
to quit this submenu.
NOTE: the fi ber optics shut off when the turbine is not used for 30
seconds (rheostat switch off).
Spray to instrument button.
It is possible to select the type of spray delivered to the instrument by
pressing the relative button.
The relative icon (
B
) appears on the console display to signal the type
of system operation.
• Icon OFF: operation with water spray + air.
• Icon ON: operation with only water spray.
NOTE: if the foot control is set to bistable operation (see paragraph
5.2.1.), it is possible to set the instrument so that water spray is always
delivered during operation.
In this case, the type of operation is indicated by the icon ( D ) on the
console display:
• Icon OFF: operation without water spray.
• Icon ON: operation with water spray.
Removable cord
The turbine has a removable cord to ease cleaning (see paragraph 5.).