VII. Maintenance
VII. Maintenance
7.1 Cleaning
7.1.1
Cleaning the instrument
Use a soft and lint-free cloth to wipe the outer surfaces
of the instrument. Take care not to scratch the touch-
panel. If contamination persists, moisten the cloth with
soap water. A special cleaning procedure is required for
the measuring cell.
7.1.2
Cleaning the measuring cell
Clean the measuring cell with water and cleaning solu-
tion only.
After each measuring cycle, go through the following
cleaning routine:
Fig. 7-1
Measuring cell and displacement piston
• Empty the measuring cell.
• Rinse it with water.
• Use a brush and cleaning solution (see chapter
4.5.3 "Cleaning solution", page 26) to clean measu-
ring cell and displacement piston.
• Rinse both with de-ionized water.
• Empty the measuring cell.
• Use a clean paper cloth to dry the outside of the
measuring cell - but NOT the inside of the cell or the
displacement piston.
Hint
Not cleaning the measuring cell after each measure-
ment will lead to measuring errors and prevent reprodu-
cibility of results.
To avoid deposits in the measuring cell and on the
replacement piston, fill the freshly cleaned cell with
water and leave overnight.
7.2 Maintenance
7.2.1
Personnel requirements
Maintenance shall exclusively be carried out by quali-
fied and skilled personnel.
7.2.2
Functional testing
To avoid measuring errors caused by faulty operation of
the instrument or contamination of the measuring cell,
functional testing can be performed by different
methods.
7.2.2.1 Basic test of PCD-05 without measuring
cell
• Turn on power switch (see Fig. 3-3, page 21).
• Switch on displacement piston drive.
- A streaming potential of 0 mV will be displayed
on the smart screen.
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...