IV. Installation
Screw in titrant tubes
Information
The titrant tubes are screwed in at first delivery. It is only
necessary to replace them in case of exchange.
There are 2 feeding and 2 dosing tubes available. FEED
marks the suction tubes, feeding the Travel titrator
module with titration liquid installed. DOSE marks the
dispensing side, transporting the titration solution into
the measuring cell.
• Screw in the dosing and feeding tubes in the Travel
titrator module as shown in figure below:
Fig. 4-7
Titrant tubes
Insert feeding tubes
• Insert the other side of the feeding tube into the tit-
rant bottle.
• Therefore insert the tubes end into the FEED
insert until the tube reaches the bottom of the tit-
rant bottle.
• When not using the screw-in titrant bottles but
others external bottles, insert the feeding tube
end into external bottle until the tube reaches the
bottom.
Insert dosing tubes
• The other side of the dosing tubes should be placed
in parking position as illustrated below.
Fig. 4-8
Dosing tubes in parking position
• In order to start titration place the dosing tube in the
tube holder as illustrated below:
Fig. 4-9
Dosing tube in titration position
Information
Before starting measurement, rinse the titrant tubes
according to chapter 4.6.4 "Rinsing the dosing tubes",
page 30.
Dosing tube
parking
position
Dosing tube
ready to titrate
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...