© BTG
Operation Manual Mütek
TM
PCD-05, September 2017
31
V. Measuring
V. Measuring
Information
The Mütek™ PCD-05 measures the streaming potential
of an aqueous sample. To detect its charge level a titra-
tion is performed until reaching a streaming potential of
0 mV. Titration may be performed manually or better
with help of an automatic titrator.
For manual operation of the Mütek™ PCD-05 Particle
Charge Detector, use a measuring pipette to transfer
titrant drop by drop into the measuring cell (see "Fig. 5-
6", page 39). The endpoint, that is the point of zero
charge, is reached when the streaming potential is at 0
mV.
5.1 Safety
rules
Be sure to follow the chapter I. "Basic safety instructi-
ons", page 7 before commissioning the Mütek™ PCD-
05.
5.2 Personnel
requirements
Only qualified and skilled personnel are allowed to ope-
rate the Mütek™ PCD-05.
5.3 Limitations
Multi user functionality is not supported
PCD-05 Smart Web application does not support multi-
user functionality. This means that several users can
connect to the instrument using the Wi-Fi but they will
not have possibility to synchronize each other's actions.
Web application will not show if someone else is opera-
ting with the instrument and how instrument is acting
depending on actions received from another user.
For example, if one user starts titration, another user will
not be able to see how its results are being built up
(titration curve).
Our recommendation is that one user at the time should
operate with the instrument using Wi-Fi.
Web cookies are not supported
PCD-05 Smart Web application does not have support
for web cookies. Which means that we do not store any
information in your web browser. Once you disconnect
from the instrument (Wi-Fi out of range, instrument
restart, browser refresh) you will enter the main page
and default states will be displayed.
For example, once you start titration button "Start Titra-
tion" will change to "Stop Titration" and you see values
of demand, potential (and pH) being updated, see
picture "PCD-05 Smart on-going titration". In the case
you do refresh of your web browser, on-going titration in
Measure tab will go back to the default state and on-
going titration is no longer displayed. You might still
hear the sound in the instrument, but Web application
shows its default state, see figure below of "PCD-05
Smart Measure Tab in default state".
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...