44
Operation Manual Mütek
TM
PCD-05, September 2017
© BTG
V. Measuring
To store the results on your mobile device push the but-
ton "Export Table", this will generate a .csv file which
can be opened in a separate program for example MS
Excel.
5.12 Methods
5.12.1 Method settings
For charge demand measurements method 1 (Method
900) and for IEP measurements method 2 (Method 901)
are preset as default method. Methods 3-9 are available
for userdefined programming.
To set/change parameters of a titration method proceed
as follows:
• Select the tab "Methods".
Information
The tab "Method" is used to program specific methods.
Selecting a method for measuring the right method
number has to be set under the tab "Measure".
Overview of functions:
Information
For fix incremental dosage (continuous), set amount
dV(min) = dV(max).
Field
Description.
Counter
Samplename
ID of the sample
StartPot
Streaming potential in mV at the start
of a measurement.
Demand
Demand of titrant which is dosed in
ml when termination criteria is rea-
ched.
pH
- EP measurement:
pH value at the start point of a mea-
surement
- IEP measurement
pH value when the mV termination
criteria is reached.
Method
1 (Method 900) and 2 (Method 901)
are preset as default methods and
not adjustable.
Methods 3-9 are available for user-
defined programming.
General
Termination
Endpoint criteria mV or pH
Pump
Toggle field for pump control:
"Automatic" selects the oppositely
charged titrant solution automati-
cally according to mV-sign (+/-).
"-" pumps from the rear pump
"+" pumps from the front pump
Type
Toggle field for measurement type:
EP charge measurement or
IEP iso electric point measurement
(pH option has to be installed)
Mode
Measuring
Mode
Fixed increment or adaptiv titration
Addition
Addition
Continuous or dynamic of titrant
Save
Termination and storing of created
method.
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...