52
Operation Manual Mütek
TM
PCD-05, September 2017
© BTG
VI. Diagnostics and troubleshooting
6.3.2
Basic faults
The following basic faults may be experienced with the Mütek™ PCD-05 also when used in combination with an exter-
nal titrator:
Problem
Cause
Solution
Heavy fluctuations of strea-
ming potential after sample
is filled in
- Excessive conductivity of the
sample
• Dilute sample.
- Insufficient sample volume in mea-
suring cell
• Weigh in once again or use
pipette.
- High solids in the sample
• Dilute the sample or centrifuge/
screen it to remove coarse par-
ticles.
Potential reverses after
reaching the endpoint
- Very slow reaction between titrant
and sample
• Titrate more slowly or perform
back titration.
• If necessary, check sample prepa-
ration for homogeneity.
At the start of titration,
potential increases instead
of decreasing
- Wrong titrant
• Select titrant with opposite charge
to that of sample.
- Adsorption equilibrium not attained
• Proceed slowly with titration.
- Sample concentration too high or
titrant concentration too low
• Dilute sample.
Potential remains essenti-
ally unchanged during titra-
tion
- Sample concentration very high
• Dilute sample or use a higher
strength titrant.
- Wet contacts on Mütek™ PCD-05
• Dry contact pins with a soft cloth.
- Wet electrodes on the outside of
the measuring cell
• Dry electrodes with a soft cloth.
- Measuring cell contaminated
• Thoroughly clean measuring cell
and displacement piston.
Sign of displayed potential
not in conformity with
sample specification
- Sample charge is pH dependent
• Check on pH dependence of
sample.
- Excessive viscosity
• Dilute sample.
- Measuring cell contaminated
• Thoroughly clean measuring cell
and displacement piston.
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...