© BTG
Operation Manual Mütek
TM
PCD-05, September 2017
49
V. Measuring
5.13.3 Service Settings
Information
With the following settings the instruments calibration
will be changed, this has a direct influence to the mea-
surement results. This setting should be changed only
from a skilled user.
Pump volume:
The pump volume can be changed between 4.5 and
5.5.
Analog input (potential):
1. Select either cell 1 or cell 2 depending which cell you
want to calibrate
2. Clean measuring cell manually
3. Fill in 10 ml 0.001 N PES Na
4. Insert measuring cell and hook piston
5. Press "Start" in the "Analog In (potential)" field
6. Wait until it becomes stable, value in the brackets
should be below 100 then press "Stop"
Mastercode:
Specific adjustment can be done, only if a code from
BTG Instruments is provided.
5.14 Help
Information
Under the tab "Help" the following can be found:
• Training and instruction movies.
• Operation Manual.