© BTG
Operation Manual Mütek
TM
PCD-05, September 2017
47
V. Measuring
5.13 Settings
5.13.1 General Settings
5.13.2 pH calibration (optional)
Information
If the Mütek™ PCD-05 is connected to an external titra-
tor system, the pH electrode has to be simultaneously
calibrated on the titrator and Mütek™ PCD-05.
For calibration, use two commercial buffer solutions,
one of them with pH 7.0 and the other with pH 4.01 or
pH 9.21.
Information
The buffers used for calibration should correspond to
the pH range of the samples.
Procedure
• Under the tab "Settings", select "Calibrate pH".
• Depending on the range of the second buffer solu-
tion you wish to use, touch the appropriate field
(acid pH 4.01 or alkaline pH 9.21).
Software version Shows the software version of the
base unit
Serial number
Serial number of the PCD-05
Titration unit
Select ml or µeq/l, measurement
results will be displayed accor-
dingly
Settling time
Alternative to start a titration
manually, it is possible to enter fix
waiting time before titration starts
automatically.
For example settling time 30 s,
after the potential is measured for
30 seconds, the titration starts.
0 s = manual start of titration
Service Settings
By entering a password 4374
further calibration can be done,
this should be done only from
advanced users, see chapter
5.13.3, page 49
Calibrate pH
Opens the "pH Calibration"
window for pH electrodes, see
chapter 5.13.2, page 47
Set Time
Will ask you to set instrument
time according to client time
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...