VII. Maintenance
7.3 Proper
handling
of the pH electrode
The pH electrode must not fall dry. Therefore always put
on the protection cap, when not using the electrode.
Alternatively use a tiny beaker with 3M KCl solution for
parking position inbetween measurements.
Always fill up the protection cap with fresh 3M KCl solu-
tion when you’ve finished operation. This will regenerate
the electrode ever time after use and will prolong the life
time of the electrode. The pH electrode will consum the
KCl ions, therefore it is needed to always use fresh KCl
solution.
Be carefull when laying the pH electrode on a table, in
order not to break the diaphragm. It is recommend
always to insert the pH electrode into a tiny beaker with
3M KCl solution for parking position inbetween measu-
rements.
Always calibrate the pH electrode (see chapter 5.13.2
"pH calibration (optional)", page 47) prior to your mea-
surements (at beginning of every working day with the
unit).
The buffer solutions supplied are meant as single use
solutions. In order to assure reproducible results always
use new/ stable buffer solutions.
7.4 Yearly LAB Check™ inspection
To guarantee reliable measurement from the start,
every Mütek™ laboratory instrument undergoes thoro-
ugh testing before leaving the factory.
BTG’s maintenance program LAB Check™ keeps your
laboratory device in perfect working order year after
year. Feel free to call in your local service team or send
in your system for factory maintenance.
To keep track of the next inspection interval, the smart
display on Mütek™ PCD-05 indicates following remin-
der messages:
- 2 weeks before the service interval is due, a service
reminder message will show up, when switching on
the Mütek™ PCD-05:
- When the service date has already been reached or
been exceeded, following message will appear,
when switching on the Mütek™ PCD-05:
A service sticker will show when the last service has
been performed and the next one should be.
PCD-05 Smart
SN 90300259 V2.5 Key ****
OK
Next service [dd.mm.yyyy]:
01.07.2018
PCD-05 Smart
SN 90300259 V2.5 Key ****
OK
Information:
Service recommended !
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...