© BTG
Operation Manual Mütek
TM
PCD-05, September 2017
51
VI. Diagnostics and troubleshooting
VI. Diagnostics and troubleshooting
6.1 Safety
instruction
If errors occur that are not listed in the table below, dis-
connect the Mütek™ PCD-05 and tag it as "defective".
Contact your local BTG division to ensure quick correc-
tive action.
6.2 Personnel
requirements
Troubleshooting may only be performed by qualified
and skilled personnel.
6.3 Table of malfunction
6.3.1
Error messages on the PCD-05 Smart
Error message
Cause
Corrective action
Warning:
No endpoint detected
- Maximum titrant volume of 20
ml reached
- Maximum titration steps of 2200
steps reached
- pH end point is reached before
0 mV is reached
• Verify correct position of anionic
and cationic titrant.
• Sample concentration too high,
dilute sample.
• Verify that titrant tubes are immer-
sed in sample.
• Rinse titrant tubes thoroughly to
remove air bubbles.
• Increment too low adjust method
Slope out of range!
The values are not saved
- pH electrode contaminated or
defective
• Clean pH electrode, if necessary
replace it.
Error:
Transfer failed!
- USB stick not identified or wri-
teable
• Unplug and replug the USB stick,
retry transfer.
• Control USB stick format.
- NTFS format does not work
• Control USB stick regarding parti-
tioning.
- An USB stick with partitioning
does not work
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...