V. Measuring
5.12.1.1 Setting of standard methods
Method 900:
Method 901:
5.12.1.2 Method Mode
• Mode: "fixed", fixed increment titration.
Depending on the time increment indicated, the Mütek
TM
PCD-05 records measured values at fixed intervals.
Addition amount has to be preset dynamic or conti-
nuous in the addition screen see chapter 5.12.1.3, page
46.
Mode: "adapt", adaptive titration.
As soon as an equilibrium is reached within a time
window, the Mütek
TM
PCD-05 continues with the next
dispensing step.
dE [mV]/ dt [s]:
Time window (dt) in which the variation of the mV signal
is below dE.
t(min) [s]/ t(max) [s]:
This display gives the minimum and maximum wait
times between two dispensing steps.
Addition amount has to be preset dynamic or continuos
in the addition screen see chapter 5.12.1.3, page 46.
Termination
mV
EP
0 mV
Pump
Automatic
Type
EP
Meas. Mode
fixed
Increment
3 s
Addition
Dynamic
dE (set)
8 mV
dV (min)
0,01 ml
dV (max)
0,1 ml
Termination
pH
EP
7,00 pH
Pump
Automatic
Type
IEP
Meas. Mode
Adaptive
dE
1,0 mV
dt
1 s
t (min)
5 s
t (max)
15 s
Addition
Dynamic
dE (set)
8 mV
dV (min)
0,005 ml
dV (max)
0,1 ml
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...