IV. Installation
4.6.3.3 Checking & setting titrant parameters
When the titrant bottles have been installed, the titrant
parameters have to be entered manually.
Information
All screens and menus are described under chapter 5.4
"Overview of Smart display Mütek™ PCD-05 Smart/
Light", page 32.
Procedure
Connect a Wi-Fi device (smartphone, tablet, or laptop)
to the PCD-05 Smart as described in chapter 5.5
"Connect a mobile device to PCD-05 Smart", page 34.
Make all titrant related settings in the "Titrator" tab:
• Content: maximum 7-letter name
• Type: anionic, cationic, acid or alkaline
• Best before: use-by-date
• Normality: any from 0,0001 - 0,01
• Factor: correction factor, default =1,000
• Batch: number from Titrant bottle
• Molarity: any from 0,001 - 1
Press "Save" to confirm your entry.
Press "Cancel" to cancel the entry, no changes will be
stored.
4.6.4
Rinsing the dosing tubes
Hint
Rinse the dosing tubes in order to assure that they are
bubble-free, before starting the measurement.
Procedure with Smart display
• Turn on the Mütek™ PCD-05 Smart
• After initialization press "Rinse"
• Both tubes will be rinsed with 10 ml
• After 10 ml it will stop automatically or it can be
stopped anytime in between by pressing "Home".
Procedure with Wi-Fi device
Connect a Wi-Fi device (smartphone, tablet, or laptop)
to the PCD-05 Smart as described in chapter 5.5
"Connect a mobile device to PCD-05 Smart", page 34.
• Enter a rinsing volume in the tab "Titrator", start rin-
sing either for pump +, -, or both.
Information
In order to return the rinsing amount back to the titrant
bottle, insert the dosing tube into the RINSE insert of
the Mütek™ PCD-05 Travel titrator module.
Hint
Assure that the correct rinsing opening is used. Other-
wise the titrant solution is contaminated and has to be
disposed!
returning
the rinsing amount
Содержание Mutek PCD-05 Smart
Страница 1: ...Operation Manual MütekTM PCD 05 Smart Light Particle Charge Detector ...
Страница 50: ...50 Operation Manual MütekTM PCD 05 September 2017 BTG V Measuring ...
Страница 58: ...58 Operation Manual MütekTM PCD 05 September 2017 BTG VIII Shutting down the system storage and disposal ...
Страница 61: ......
Страница 62: ...BTG Instruments AB P O Box 602 SE 661 29 Säffle Sweden Phone 46 533 426 00 www btg com ...