2.3
Temperature control system
LCDP shut-down thermal snap-switch
Opening temperature
60
±
3
o
C
Closing temperature
50
±
4
o
C
Thermocouples
K type class 1
LCMB shut-down thermal snap-switch
Opening temperature
120
±
3
o
C
Closing temperature
110
±
4
o
C
LCDP and LCMB motor-protection thermistors
Type
Positive temperature coefficient
Reference temperature
150
o
C
Compliant with
IEC 34-11 (BS4999 part III)
Water flow-switch
Closed when flow
≥
1.5 l.min
-1
2.4
Cooling-water supply
Note:
Use treated water or non-corrosive industrial water to cool the iL system.
Maximum supply pressure
100 psig (6.9 bar, 6.9 x 10
5
Pa)
Water supply temperature
10 to 30°C
Typical pressure differential
across supply and return *
iL70
15 psi (1.03 bar, 1.03 x 10
5
Pa)
iL600
15 psi (1.03 bar, 1.03 x 10
5
Pa)
Minimum required flow rate for
reliable system operation
iL70
1.5 l.min
-1
iL600
2.5 l.min
-1
Typical heat removed from iL system
iL70
1.2 kW
iL600
1.8 kW
Maximum particle size in supply
0.03 mm
2
Water supply quality
Acidity
pH 6.5 to 8.0
Hardness
< 100 p.p.m
Resistivity
> 1 k
Ω
.cm
-1
Solids (turbidity)
< 100 p.p.m
* With cooling-water supply temperature of 20
o
C and flow rate of 3 l.min
-1
.
iL Dry Pumping Systems
2-9