![BOC Edwards iL600 Скачать руководство пользователя страница 102](http://html1.mh-extra.com/html/boc-edwards/il600/il600_instruction-manual_2781982102.webp)
5
OPERATION
5.1
Introduction
The following sections describe the use of the Pump Display Module to operate the iL system.
For a description of the use of an Interface Module or Communications Module accessory to
operate the iL system, refer to the instruction manual supplied with the Module. For a
description of the operation of the load-lock pump (if you have connected the iL system to the
load-lock pump through an Interface Module), refer to Section 5.14.
The following sections summarise the use of the menus in the Pump Display Module. We refer
to menus such as 'the Setup/Service/Manual menu'. Refer to Figure 1-7 or 4-1: this means that
you:
•
Press the Setup button (7) to select the Setup menu.
•
Use the up and down buttons (5, 3) to show the Service option and press ENTER (6) to select
the Service menu.
•
Use the up and down buttons (5, 3) to show the Manual option and press ENTER (6) to select
the Manual menu.
We recommend that you fully understand the use of the Pump Display Module menus and
buttons before you operate the iL system. Refer to Section 4 which fully defines the menus, the
use of the buttons and the display formats.
Note that you must switch on the electrical supply to the iL system before you can use the Pump
Display Module.
5.2
Change the display format
5.2.1
Change the normal display
During normal operation (or whenever the Normal menu is selected), the Pump Display Module
will show the first page of the normal display. The normal display has two pages and each page
can show the readings of one or more sensors or internal clocks or counters. As supplied, the
information shown on the first page is:
•
LCDP pump power consumption
•
LCDP pump-body temperature
If you press the up or down button, the Pump Display Module will then show the second page of
normal display; as supplied, the information shown on the second page is:
•
LCMB pump power consumption
("<NP>" shown for iL70 systems: see Section 4.4)
•
LCMB pump temperature
If required, you can use the Setup/Normal display menu to change the information shown on
the two normal display pages.
iL Dry Pumping Systems
5-1
Содержание iL600
Страница 18: ...iL Dry Pumping Systems 1 9 Figure 1 3 Schematic diagram of the electrical system...
Страница 21: ...1 12 iL Dry Pumping Systems Figure 1 4 Services and electrical connections on the rear of the iL system...
Страница 23: ...1 14 iL Dry Pumping Systems Figure 1 5 Electrical components shown without enclosure for clarity...
Страница 71: ...3 28 iL Dry Pumping Systems...