![BOC Edwards iL600 Скачать руководство пользователя страница 119](http://html1.mh-extra.com/html/boc-edwards/il600/il600_instruction-manual_2781982119.webp)
6.6
Check the LCMB pump oil-level (iL600 only)
WARNING
If you have pumped dangerous substances, ensure that you purge the iL system
(refer to Section 6.1.2) before you remove the oil filler-plug.
WARNING
Do not remove the oil filler-plug while the pump is operating. If you do, hot oil may be
ejected from the pump.
CAUTION
Do not check the oil-level when the iL system is operating. The correct oil-level
may not be shown when the pump is operating.
CAUTION
Ensure that the oil-level in the LCMB pump is correct. If the pump oil-level is incorrect,
pump performance may be affected and the pump may be damaged.
Note:
If there is a significant loss of oil from the pump, the shaft-seals or the spinning filters may have
failed. You cannot replace these items. Contact your supplier or BOC Edwards for advice.
You must check the oil-level every three months; use the following procedure:
1.
Purge the system (refer to Section 6.1.2), then shut down the iL system (refer to Section 5.8)
and allow it to cool.
2.
Remove the right-hand side LCMB enclosure panel from the iL system: refer to Section 3.4.
3.
Refer to Figure 1-1. Check the oil-level against the MAX and MIN mark (see detail B) on the
bezel of the LCMB oil-level sight-glass (4):
•
If the oil-level is above the MAX mark, drain excess oil from the pump as described in
Section 6.9 until the oil-level is correct.
•
If the oil-level is below the MIN mark, fill the pump with oil until the oil-level is correct,
as described in Section 6.9.
4.
Refit the right-hand side LCMB enclosure panel: refer to Section 3.17.
6-8
iL Dry Pumping Systems
Содержание iL600
Страница 18: ...iL Dry Pumping Systems 1 9 Figure 1 3 Schematic diagram of the electrical system...
Страница 21: ...1 12 iL Dry Pumping Systems Figure 1 4 Services and electrical connections on the rear of the iL system...
Страница 23: ...1 14 iL Dry Pumping Systems Figure 1 5 Electrical components shown without enclosure for clarity...
Страница 71: ...3 28 iL Dry Pumping Systems...