6.8
Change the LCDP pump oil
WARNING
If you have pumped dangerous substances, ensure that you purge the iL system
(refer to Section 6.1.2) before you remove the oil filler-plug.
WARNING
Do not remove the oil filler-plug while the pump is operating. If you do, hot oil may be
ejected from the pump.
CAUTION
Ensure that the oil-level in the LCDP pump is correct. If the pump oil-level is incorrect,
pump performance may be affected and the pump may be damaged.
1.
Purge the system (refer to Section 6.1.2), then shut down the iL system (refer to Section 5.8)
and allow it to cool.
2.
Remove the right-hand side LCDP enclosure panel from the iL system: refer to Section 3.4.
3.
Refer to Figure 1-1. Remove the LCDP oil filler-plug (7), then use a suitable pump or syringe
to suck the oil out of the pump.
4.
Dispose of the old 'O' ring on the oil filler-plug (7) and replace it with a new 'O' ring (available
as a spare: refer to Section 8.3).
5.
Fill the pump through the filler hole with the correct grade and quantity of oil. Allow the oil
to drain into the pump, then check the oil-level on the sight-glass: refer to Section 6.7.
6.
Ensure that the new 'O' ring is correctly positioned on the oil filler-plug (7), then refit the
filler-plug to the LCDP pump.
7.
Refit the right-hand side LCDP enclosure panel: refer to Section 3.17.
6-10
iL Dry Pumping Systems