Error Codes
|
193
Synergy HT Operator's Manual
Code
Description and Probable Causes:
2500
X-axis went by flash location too soon during sweep mode
This error indicates that during sweep mode, the unit calculated the exact time for the flash to
occur and the flash failed to flash at that time.
Probable Causes:
•
Defective motor driver PCB.
•
One of the carrier axis motors is defective.
2505
Monochromator position more than 1 full step past flash point
This error indicates that during a spectral scan, the monochromator went more than 1 step past
the flash point.
Probable Causes:
•
Monochromator motor is defective.
•
Motor power supply PCB lost or added counts from monochromator motor.
2600
Physical limit exceeded for area scan request
This error can occur only during computer control, indicating that the area scan requested is too
large for the unit. The X-axis tripped the XY sensor.
The Corning
®
Costar 12 well plate (part number 430345) has a width dimension that is causing a
2600 error with Synergy. Simply changing the width to 84320 will alleviate the error message and
will not affect the results.
In KC4 v3.4 r16, the width was changed to 84320 and the 2600 error does not appear when
running a protocol that specifies the Corning Costar 12 Well plate type.
280x
<Motor> currently in use
This error indicates that the <motor> is not available for this model or already has a task
assigned to it. At the beginning of the motor_setup function, the basecode checks to see if the
motor is currently in use or is not available.
This error can occur with any motor request. See table below for errors.
Error
Motor
2800
X-axis
2801
Y-axis
2802
EX motor
2803
EM motor
2804
Order-sorting filter wheel
2805
Monochromator
2806
Probe Height
2807
Probe changer
2808
Syringe 1
2809
Syringe 2
Probable Causes:
•
User selected the wrong model in the controlling software.
•
The incorrect basecode was downloaded to the instrument.
2900
Fluorescence Lamp off voltage out of range (0 – 730)
This error indicates that during a self-test, the lamp did not indicate that it was off.
Probable Causes:
•
The sense resistor is damaged on the 7090402 PCB.
•
Cable between the lamp PCB and motor power supply PCB is defective.
•
Motor power supply PCB is defective.
Содержание Synergy HT
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Страница 2: ......
Страница 22: ...xx Preface BioTek Instruments ...
Страница 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Страница 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Страница 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...