Flushing/Purging the Fluid Path
|
129
Synergy HT Operator's Manual
Flushing/Purging the Fluid Path
Applies only to Synergy™ HT models with injectors.
At the end of each day that the dispense module is in use,
flush
the fluid path using
Gen5’s or KC4’s priming utility. Leave the fluid to soak overnight or over a weekend, and
then
purge
the fluid before using the instrument again.
Note:
This flushing and purging routine is also recommended before disconnecting the
outlet tubes from the rear of the reader, and before decontamination to remove any assay
residue prior to applying isopropyl alcohol or sodium hypochlorite.
To
flush
the fluid path:
1.
Fill two supply bottles with deionized or distilled water. Insert the supply
(inlet) tubes into the bottles.
2.
Place the priming plate on the carrier.
3.
Gen5:
Select
System|Reader Control|Synergy(Com<#>)
.
KC4:
Select
System|Reader Control
.
4.
Click the
Dispenser
tab and select
Dispenser 1
.
5.
Set the
Volume
to 5000 L. Keep the default prime
Rate
.
6.
Click
Prime
to start the process. When the process is complete, carefully
remove the priming plate from the carrier and empty it.
7.
Repeat the process for
Dispenser 2
.
Leave the water in the system overnight or until the instrument will be used
again. Purge the fluid from the system (see below) and then prime with the
dispense reagent before running an assay.
To
purge
the fluid from the system:
1.
Place the inlet tubes in empty supply bottles or a beaker.
2.
Gen5:
Select
System|Reader Control|Synergy(Com<#>)
.
KC4:
Select
System|Reader Control
.
3.
Click the
Dispenser
tab and select
Dispenser 1
.
4.
Set the
Volume
to 2000 L.
5.
Click
Purge
to start the process.
6.
When the purge is complete, repeat the process for
Dispenser 2
.
After purging the system, you may wish to run a quick Dispense
protocol to visually verify the dispense accuracy. See the next page
for instructions for creating the protocol.
Содержание Synergy HT
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Страница 2: ......
Страница 22: ...xx Preface BioTek Instruments ...
Страница 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Страница 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Страница 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...