Cleaning the Internal Components
|
147
Synergy HT Operator's Manual
9.
Use a 3/32” Allen wrench to remove the two shoulder screws securing the
top probe hanger. Remove the screws and set them aside.
10.
Drop the top probe hanger down and slide to the left to remove it. Turn the
hanger upside down to clean the absorbance lens (see instructions on the
next page).
Do not touch the lens with your fingers!
Inspect the block
for spills or other contamination.
Carefully
clean with mild detergent if
necessary.
Shoulder
screws
Absorbance
Lens
Probe hanger
(bottom view)