Status String Format
|
205
Synergy HT Operator's Manual
Dispense/Read Runtime/Definition Error Test Codes (lowest two digits in returned error code)
SAMPLE START LATE
“
0
”
individual sample missed its scheduled start time
FILTERSETS OUT OF RANGE
“
01
”
more than two filters selected for disp/read method
FILTER SWITCHING OVERLAP
“
02
”
filter bandpass overlap condition exists in the
intermediate switching position for multi filterset
reads.
FILTERS NOT ADJACENT
“
03
”
ex or em filters not adjacent for multi filterset
reads
EX FILTER PLUG POSITION
“
04
”
filter plug(s) not positioned next to ex filter(s)
for light shutter feature
INVALID MODE
“
05
”
invalid processing mode received
DISP READ DEF NOT SET
“
06
”
dispense/read protocol parameters not set or not
valid
FEATURE NOT AVAILABLE
“
07
”
disp/read feature not enabled by dip switch
INVALID EX WAVE
“
08
”
excitation wavelengths must be the same for TR assay
TIP PRIME TUB VOL
“09”
total tip prime vols in assay larger than tub
capacity
NUM EVENTS
“
0B
”
total number of events defined invalid
INVALID EVENT
“
C
”
invalid event type received
INVALID GEOMETRY
“
0D
”
wells spaced too close for dispense operation
PLATE START LATE
“
0E
”
plate mode event missed its scheduled start time
INVALID TIP PRIME
“
0F
”
tip prime conflicts with tip prime rules
NUM READ EVENTS
“
10
”
number of read events defined invalid
READ START TIME
“
11
”
read event start time out of range
NUM SAMPLES
“
12
”
number of samples out of range
SAMPLING INTERVAL
“
13
”
sampling interval out of range
NUM PLATE READS
“
14
”
number of plate mode kinetic reads out of range
PLATE MODE READ TIME
“
15
”
plate mode kinetic interval time too short
READ START LATE
“
16
”
read event missed its scheduled start time
NUM DISP EVENTS
“
21
”
number of dispense events defined invalid
TIP PRIME VOLUME
“
22
”
tip prime volume out of range
DISP VOLUME
“
23
”
volume requested out of range
DISP RATE
“24
”
dispense rate requested out of range
DISP START TIME
“
25
”
dispense start time/kinetic interval out of range
DISP INTERVAL
“
26
”
dispense kinetic interval out of range
DISPENSER NOT ATTACHED
“
28
”
dispenser module not attached
DISPENSER NOT INITIALIZED
“
29
”
dispenser has not been initialized
DISPENSER NOT PRIMED
“
2A
”
dispenser has not been primed
INVALID DISPENSER OPERATION
“
2B
”
invalid dispenser operation type requested
DISP START LATE
“
2C
”
dispense event missed its scheduled start time
NUM SHAKE EVENTS
“
30
”
number of shake events defined invalid
SHAKE START TIME
“
31
”
shake start time out of range
READ EVENT EXPECTED
“
32
”
read event must follow shake event with repeat
SHAKE TIME
“
33
”
shake time out of range
Содержание Synergy HT
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Страница 2: ......
Страница 22: ...xx Preface BioTek Instruments ...
Страница 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Страница 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Страница 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...