Hazards and Precautions
|
xv
Synergy HT Operator’s Manual
Warning! Software Quality Control.
The operator must follow the manufacturer’s
assay package insert when modifying software parameters and establishing reading
methods.
Failure to conduct quality control checks could result in erroneous
test data
.
Warning! Pinch Hazard.
Some areas of the Dispense Module can present pinch
hazards when the instrument is operating. These areas are marked with the symbol
shown here. Keep hands/fingers clear of these areas when the instrument is
operating.
Precautions
The following precautions are provided to help avoid damage to the instrument:
Caution: Service.
The Synergy™ HT should be serviced by BioTek authorized
service personnel. Only qualified technical personnel should perform troubleshooting
and service procedures on internal components.
Caution: Environmental Conditions.
Do not expose the system to temperature
extremes. For proper operation, ambient temperatures should remain between
18
°
-40
°
C. Performance may be adversely affected if temperatures fluctuate above or
below this range. Storage temperature limits are broader.
Caution: Sodium Hypochlorite.
Do not expose any part of the instrument to the
recommended diluted sodium hypochlorite solution (bleach) for more than 20
minutes. Prolonged contact may damage the instrument surfaces. Be certain to rinse
and thoroughly wipe all surfaces.
Caution: Power Supply.
Only use the power supply shipped with the instrument.
Operate this power supply within the range of line voltages listed on it.
Caution: Shipping Panel and Carrier Shipping Screw.
The shipping panel and
carrier shipping screw must be removed before operating the reader. They must be
reinstalled before repackaging the reader for shipment. See
Chapter 2, Installation
.
Caution: Disposal.
This instrument contains printed circuit boards and wiring with
lead solder. Dispose of the instrument according to Directive 2002/96/EC, “on waste
electrical and electronic equipment (WEEE).”
Caution: Electromagnetic Environment.
Per IEC 61326-2-6 it is the user’s
responsibility to ensure that a compatible electromagnetic environment for this
instrument is provided and maintained in order that the device will perform as
intended.
Caution: Electromagnetic Compatibility.
Do not use this device in close
proximity to sources of strong electromagnetic radiation (e.g. unshielded intentional
RF sources), as these may interfere with the proper operation.
Caution: Warranty.
Failure to follow preventive maintenance protocols may
void
the warranty
.
See
Chapter 5, Preventive Maintenance
.
Содержание Synergy HT
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Страница 2: ......
Страница 22: ...xx Preface BioTek Instruments ...
Страница 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Страница 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Страница 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...