IQ/OQ/PQ
|
63
Synergy HT Operator's Manual
Operational Qualification
confirms that the equipment operates according to
specification initially and over time.
•
The recommended OQ procedure consists of performing the System Test,
Absorbance Plate Test, a series of Liquid Tests, and, if the external dispense
module is used, the Dispense Accuracy and Precision Tests.
•
Your facility’s operating policies may also require that you perform an actual assay
prior to accepting the reader for routine use. If this is the case, you should not use
the data obtained from the first assay run on the reader until you have confirmed
that the package insert criteria have been met.
•
The OQ procedure should be performed
initially
(before first use) and then
routinely; the recommended interval is
annually
. It should also be performed after
any major repair or upgrade to the hardware or software.
•
Although out-of-tolerance failures will be detected by the OQ tests, results should
be compared with those from the routine Performance Qualification tests and
previous OQ tests to monitor for trends.
•
The successful completion of the OQ procedure, in combination with results that
are comparable to previous PQ and OQ tests, confirms that the equipment is
operating according to specification initially and over time.
Performance Qualification
confirms that the reader consistently meets the
requirements of the tests performed at your laboratory.
•
The recommended PQ procedure consists of performing the System Test,
Absorbance Plate Test, a series of Liquid Tests, and, if the external dispense
module is used, the Dispense Accuracy and Precision Tests.
•
Your facility’s operating policies may also require that you routinely perform an
actual assay, to confirm that the reader will consistently give adequate results for
the assays to be run on it.
•
These tests should be performed routinely; the recommended interval is
monthly
or
quarterly
, depending on the test. This frequency may be adjusted depending on the
trends observed over time.
•
The successful completion of the PQ procedure confirms that the equipment is
performing consistently under normal operating conditions.
Содержание Synergy HT
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Страница 2: ......
Страница 22: ...xx Preface BioTek Instruments ...
Страница 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Страница 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Страница 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...