156
|
Appendix A: Decontamination
BioTek Instruments, Inc.
Procedure for Models without Injectors
The
sodium hypochlorite (bleach)
solution is caustic;
wear gloves and eye protection when handling the solution.
Do not immerse the instrument, spray it with liquid, or use a
“wet” cloth. Do not allow the cleaning solution to run into the
interior of the instrument. If this happens, contact the BioTek
Service Department.
L
Important!
Turn off and unplug the instrument for all
decontamination and cleaning operations.
1.
Turn off and unplug the instrument.
2.
Prepare an aqueous solution of 0.50% sodium hypochlorite (bleach). As an
alternative, 70% isopropyl alcohol may be used if the effects of bleach are a
concern.
Be sure to check the percent NaClO of the bleach you are using; this
information is printed on the side of the bottle. Commercial bleach is
typically 10.0% NaClO; if this is the case, prepare a 1:20 dilution.
Household bleach is typically 5.0% NaClO; if this is the case, prepare a
1:10 dilution.
3.
Moisten a cloth with the bleach solution or alcohol.
Do not soak the cloth.
4.
Manually open the plate carrier door; slide out the plate carrier.
5.
Wipe the plate carrier and all exposed surfaces of the instrument.
6.
Wait 20 minutes. Moisten a cloth with deionized (DI) or distilled water and
wipe all surfaces of the instrument that have been cleaned with the bleach
solution or alcohol.
7.
Use a clean, dry cloth to dry all wet surfaces.
8.
Reassemble the instrument as necessary.
9.
Discard the used gloves and cloths using a Biohazard trash bag and an
approved Biohazard container.
Содержание Synergy HT
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Страница 2: ......
Страница 22: ...xx Preface BioTek Instruments ...
Страница 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Страница 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Страница 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...