Appendix A: Decontamination
BioTek Instruments, Inc.
6.
Remove the external tubing and the syringes from the dispense module and
soak them in the bleach or alcohol solution.
Wait for 20 minutes.
To remove the syringes:
Pull down the syringe bracket until it stops.
Remove the metal thumbscrew from underneath the bracket. Unscrew
the top of the syringe from the bottom of the syringe drive. Gently
remove the syringe and store it in its original packaging (see
Chapter
2, Installation
).
7.
Moisten a cloth with DI or distilled water and wipe all surfaces that have been
cleaned with the bleach solution or alcohol.
8.
Rinse all tubing and the syringes with DI water.
9.
Use a clean, dry cloth to dry all wet surfaces on the instrument and the
Dispense module.
10.
Reassemble the instrument and dispense module as necessary.
11.
Discard the used gloves and cloths using a Biohazard trash bag and an
approved Biohazard container.