Cleaning Exposed Surfaces
|
127
Synergy HT Operator's Manual
Cleaning Exposed Surfaces
Important!
Turn off and unplug the instrument for all
cleaning operations.
Important!
Do not immerse the instrument, spray it with
liquid, or use a “wet” cloth. Do not allow the cleaning
solution to run into the interior of the instrument. If this
happens, contact BioTek’s Service Department.
Exposed surfaces may be cleaned (not decontaminated) with a cloth moistened
(not soaked) with water or water and a mild detergent. You’ll need:
•
Deionized or distilled water
•
Clean, lint-free cotton cloths
•
Mild detergent (optional)
To clean the exposed surfaces:
1.
Turn off and unplug the instrument.
2.
Moisten a clean cotton cloth with water, or with water and mild detergent.
Do not soak the cloth
.
3.
Wipe the plate carrier and all exposed surfaces of the instrument.
4.
Wipe all exposed surfaces of the dispense module (if used).
5.
If detergent was used, wipe all surfaces with a cloth moistened with water.
6.
Use a clean, dry cloth to dry all wet surfaces.
7.
Reassemble the instrument as necessary.
Models with injectors:
If the Tip Priming Trough overflows, wipe the
carrier and the surface beneath the carrier with a dry cotton cloth. If
overflow is significant, you may have to remove the shroud of the
instrument to better access the surface beneath the carrier.
See page 135 for instructions on removing the shroud.
See page 149 for instructions for cleaning the surface beneath the
carrier.
Содержание Synergy HT
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Страница 2: ......
Страница 22: ...xx Preface BioTek Instruments ...
Страница 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Страница 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Страница 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...