Error Codes
|
171
Synergy HT Operator's Manual
Code
Description and Probable Causes:
0300
Saturation transition failed in the X-axis movement
(light beam never found)
This error indicates that during the X-axis movement, the light beam (saturation) transition (max
light to no light) was never found during autocalibration.
Probable Causes:
Absorbance mode:
•
The rail, nylon bushings, or bearings are dirty, causing the carrier to jam.
•
The Autocal jig is not in the carrier.
•
Order-sorting filter wheel is jammed and is not able to turn to the open hole.
Fluorescence mode:
•
The rail, nylon bushings, or bearings are dirty.
•
The Autocal jig is not in the carrier, or mirrors are facing the wrong direction.
0301
Saturation transition failed in the Y-axis movement
(light beam never found)
This error indicates that during the Y-axis movement, the light beam (saturation) transition (max
light to no light) was never found.
Probable Causes:
Absorbance mode:
•
The rail, nylon bushings, or bearings are dirty, causing the carrier to jam.
•
The Autocal jig is not in the carrier.
•
Order-sorting filter wheel is jammed and is not able to turn to the open hole.
Fluorescence mode:
•
The rail, nylon bushings, or bearings are dirty.
•
The Autocal jig is not in the carrier, or mirrors are facing the wrong direction.
0302/
0303
EX/EM filter wheel did not home
(This error code indicates older basecode. Recommend upgrading.) This error indicates the filter
wheel did not rotate to activate the Hall Effect home sensor.
Probable Causes:
•
The Filter Wheel is removed when the instrument is powered on and the wheel positioning
is changed.
•
The filter wheel is jammed. Reseating the filter wheel may correct this.
0304
Monochromator order-sorting filter wheel did not home
This error is caused when the Order-sorting Filter Wheel cannot rotate to the home sensor
position.
Probable Causes:
•
The motor or motor driver circuit is defective.
•
The home sensor or sensor circuit is defective.
Содержание Synergy HT
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HT ...
Страница 2: ......
Страница 22: ...xx Preface BioTek Instruments ...
Страница 142: ...120 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 144: ...122 Chapter 4 Instrument Qualification BioTek Instruments Inc ...
Страница 174: ...152 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 186: ...164 Appendix B Computer Control BioTek Instruments Inc ...
Страница 236: ...214 Appendix D Specifications BioTek Instruments Inc ...
Страница 240: ...218 Appendix E Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...