411829-L, Description and Specifications
April 2012
Page 1-17
1
1.3 Specifications
1.3.1 General System Specifications
Power
Input Voltage
100-240 VAC, 50/60 Hz
Fuses
4.0 TA/250 V, P/N 411916 (2 required)
Max Power Consumed
200 VA
Certification
Certificates
CE, UL, CAN/CSA
Directives
LVD, Directive 2006/95/EC
EMC, Directive 2004/108/EC
WEEE, Directive 2002/96/EC
Standards
EN 61010-1: 2001
EN 1127-1:2007
EN 61326-1:2006
CAN/CSA-C22.2 No. 61010-1-04
UL 61010-1:2004
Fluid Path (Wetted) Materials
The wetted parts of the system contain stainless steel, FFKM, PEEK,
PP, PE, FEP, PTFE, silica, carbon fiber filled PTFE,
and
fused silica, which are not affected by common chromatographic solvents.
The Isolera LS system is delivered with two kinds of peristaltic pump tubes. Ensure that the used peristaltic pump
tube is compatible with the solvents and sample to be used; see section 1.3.4 on page 1-20.
Dimensions and Weights
Dimensions
Width: 355 mm (14") or, when using a system with two collection trays, 577 mm
(22.7");
Height: 596 mm (23.5");
Depth: 497 mm (19.6")
At least three centimeters (one inch) of space should be maintained between the
system’s rear panel and other objects to allow proper ventilation.
Weights
30-35 kg (66-77 lbs) depending on system configuration
Environmental
Location
A laboratory fume hood with capacity to handle leakage of solvents. If using an Isolera
LS Funnel Rack Kit, a walk-in fume hood must be used.
Temperature
15°C to 32°C (59°F to 89.6°F)
Humidity
20% to 95% at room temperature
Interfaces
LAN port, two USB ports (2.0), mouse port, VGA port, external detector port, and leak detector port