411829-L, System Settings
April 2012
Page 3-14
3
3.7.9 Release Stuck Check Valves
Low or inconsistent flow delivery volume and/or superimposed periodic UV or UV-VIS signals can be
signs of sticking check valves. Use the following procedure to release stuck check valves.
1. Select the
Maintenance
tab in the right-hand panel.
2. Press
Release Check Valves
in the
Valves
field. Read and follow the instructions that appear
on the screen.
3.7.10 View and Reset the Usage Statistics
Both the number of runs performed on the system since it was installed and the number of runs
performed since the last reset are displayed in the
Usage
field at the
Maintenance
tab.
To reset the usage statistics:
1. Select the
Maintenance
tab in the right-hand panel.
2. Press
Reset
in the
Usage
field. The
Reset Performed Runs
dialog opens.
3. To confirm reset, press
Yes
.