411829-L, Operation
April 2012
Page 4-18
4
To reuse the peristaltic pump tube used in the previous run (not recommended – see the note
above), flush and fill the tubing with a solvent suitable for the next purification:
1. Place the pump’s inlet line into the solvent of your choice.
2. Place the pump’s outlet line into a waste reservoir.
Figure 4-13. Sample Loading Pump
3. Select the
Solvents
tab in the right-hand panel (in the software) and then the
Sample
Loading Pump
tab.
4. To start the pump, press
Start
. Adjust the flow rate using the
Speed
slider.
5. When you are done, press
Stop
.
Figure 4-14. Sample Loading Pump Tab
6. Connect the pump’s outlet line to the cartridge’s inlet.
7. Place the pump’s inlet line into the sample.
Outlet line
Inlet line