411829-L, Collection and Fractionation
April 2012
Page A-4
A
System Parameters
The system parameters are defined for all methods and can be changed (in the System mode) only
by a user with system owner privilege. The following system parameters are available:
Start Slope
The value of the collect start slope. The range is [Off | 1 to 5000 mAU/CV]. (Only
available with the UV1 and UV1+UV2 detection modes.)
Shoulder Slope
The value of the fractionation shoulder slope. The range is [Off | 1 to 5000 mAU/CV].
(Only available with the UV1 and UV1+UV2 detection modes.)
Table A-3: System Parameters
Parameter
Description
Start Slope Enable
The level where the parameter Start Slope is enabled. Below this level the Start
Slope parameter is ignored. The range is [0 to 6000 mAU]. The default value is
30
mAU.
Shoulder/Valley Slope
Disable
The level where the parameters Shoulder Slope and Valley Slope are disabled.
Above this level the Shoulder Slope and Valley Slope parameters are ignored. The
range is [0 to 6000 mAU]. The default value is 5500 mAU.
Valley Slope
The value of the fractionation valley slope. The range is [Off | 1 to 5000 mAU/CV].
The default value is 150 mAU/CV.
Table A-2: Method Parameters
Parameter
Description