411829-L, Description and Specifications
April 2012
Page 1-20
1
1.3.4 Chemical Resistance of the Peristaltic Pump Tube (Isolera LS)
The Isolera LS system is delivered with two kinds of peristaltic pump tubes. Ensure that the used
pump tube is compatible with the solvents and sample to be used.
NOTE
The peristaltic pump tube has limited use. To avoid the risk of cross-
contamination and decreased pump performance, it is recommended that
the peristaltic pump tube is discarded after each run. If you still want to
reuse the peristaltic pump tube, clean the tube as soon as possible as
described in section 5.8.1 on page 5-5.
Limited: The peristaltic pump tube must be discarded after each run.
The chemical resistance was tested in room temperature as follows: The peristaltic tube was
1) weighed, 2) submerged in solvent for three hours, 3) dried and visually inspected, 4) submerged for
another three hours, 5) dried, visually inspected, and weighed, and 6) dried for 24 hours and weighed.
Solvent
PharMed,
P/N 412480
Fluran,
P/N 412481
Solvent
PharMed,
P/N 412480
Fluran,
P/N 412481
Acetetic acid
Yes
No
n-Heptane
Limited
Yes
Acetone
Limited
No
n-Hexane
No
Yes
Acetonitrile
Yes
No
Isopropanol
Limited
Yes
Ammonia
Yes
Yes
Methanol
Yes
Yes
Dichloromethane
No
Yes
Pyridine
Limited
Limited
Ethanol
Yes
Yes
Toluene
No
Yes
Ethyl acetate
Limited
No
Triethylamine
No
Yes
Formic acid
Yes
Yes
Water
Yes
Yes