44 / 120
E3KOM-7.00
CardioTouch3000 Operation Manual
5) LCD Monitor display
Monitor mode display shows the currently measured electrocardiogram signal in real
time. The monitor mode display function is used for the following
two purposes: One is to
check whether the signals from all channels are coming out properly; the other one is to
see the status of the electrocardiogram of the patient for an
extended time.
The printing speed can be selected to 5mm/s, 12.5 mm/s, 25 mm/s, 50mm/s and 100
mm/s, and the sensitivity can be selected to 2.5 mm/mV, 5 mm/mV, 10 mm/mV, 20
mm/mV, and Auto.
Displayed heart rate is the average of 4 heartbeats
NOTE
The heart rate detection range is 30 to 300 bpm, and the error range is
±
2.
6) Rhythm Mode Printing
In rhythm mode printing, the measured ECG signals would be printed in real-time.
Rhythm mode printing allows operators to perform the following
two functions: one is to
verify that signals in all channels are active before recording ECGs, and the other is to print
the ECGs rhythm status of the patient for a longer duration.
Printing speed can be set to 5mm/s, 12.5 mm/s, 25mm/s, 50mm/s, 100mm/s and signal
level can be set to 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, and Auto.
When pressing the ‗RHYTHM‘ key on the control panel, you will see the message as
shown below and printing will start in accordance with the speed you set up.
Содержание CardioTouch3000
Страница 1: ...CardioTouch3000 Operation Manual Ver 7 00 2019 05 23 www ebionet com...
Страница 48: ...48 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 49: ...49 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 50: ...50 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 51: ...51 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 12 CH Record Report Form BEAT REPORT TEXT...