If the recorded EKG signal contains a lot of noise, Low Pass Filter (LPF) may help to clear it.
For Low Pass Filter, 40Hz, 100Hz, 150Hz and OFF option is offered. Here, 40Hz option means
the filter will remove any unwanted signal with frequencies above 40Hz. Low Pass Filter
values will appear on the bottom of the printout form; if LPF setting is OFF then as LPF: Off; if
the LPF setting is 40Hz then as LPF: 40Hz; if LPF setting is 100Hz then as LPF: 100Hz; if LPF
setting is 150Hz, then as LPF: 150Hz.
Select 'ON' when applying ‗EMG‘ filter; otherwise, select 'OFF.' Setup values will be shown
on the bottom of the printout form as ‗EMG:On‘ or ‗EMG:Off‘ when applying the filter.