CardioTouch3000 Operation Manual
Countermeasures for Bad Lead Connection
After turning on the power switch, enter the ECG test to check waveforms of all leads.
On this menu, operators can check waveforms of all leads and their noise, and may see
―Lead Fault‖ message displayed on the LCD screen.
NOTE
Message will be displayed when Lead fault is set to ON in System Setup, and will
NOT be displayed when set to OFF..
The following two cases could cause a lead fault error.
- First case, leads are detached from the patient‘s body. In this case, re-attach leads
normally.
- Second case, signal conductivity between leads and the patient‘s body is low. In this
case, replace the electrodes.
If the two above cases have been performed and the operator is still experiencing
noise or lead fault conditions, the patient cable may be faulty. Please contact the Bionet
service center.
Lead Fault! All Leads.